Web.Requests
Yodify DEC Brand Library
9092/9092C Chlorine/pH Analyzer

9092/9092C Chlorine/pH Analyzer

Brand: Waltron Bull & Roberts, LLC

Introducing the new Waltron 9092/9092C Chlorine Analyzer for automatic, continuous measurement of free chlorine, active chlorine, ozone, chlorine dioxide and bromine in aqueous media. 

Measurement is based on a microprocessor based potentiostatic three electrode system. The Waltron 9092/9092C series instruments are used for industrial process monitoring and process control. 

Applications include drinking water treatment, cooling water treatment, water management in swimming pools, or the control of residual concentration Cl2 of chlorine treated waste water. 

The potentiostatic technology allows for a much faster response time than membrane or amperometric systems, ensuring short start-up times, easy operation, and low maintenance.
Features
  • Portable system - Model 9092C available
  • Automatic calibration option
  • Wide range analysis: 0 to 1,000 μg/L, 0 to 10 mg/L
  • Automatic temperature and flow compensation
  • Fast response time (t90 = 30 sec)
  • No zero point adjustment required
  • Withstands pressures up to 10 bar (145 psi)
  • Analog and digital interface; data logging option
  • Minimal maintenance required
  • Automatic pH compensation for free chlorine reading

Specifications

Accuracy

  • ± 3 % of Reading

Ambient Temperature Range

  • 0° to 55° C (32° to 131° F)

Calibration

  • Automatic Calibration
  • Manual

Communication

  • Serial Interface RS-232

Conductivity

  • Sample Conductivity: ≥ 10 μS/cm, Salt Cell is Required if Conductivity is Less

Environmental Protection

  • IP54

Flow Rate

  • 3 to 18 LPH

Materials of Construction

  • Counter Electrode: Stainless Steel 1.4571 (314)
  • Measuring Electrode: Gold
  • Reference Electrode: Silver Chloride (Ag/AgCl) in Saturated KCl-Solution

Maximum System Pressure

  • psig: 145 (kPa: 1,000, bar: 10)

Output

  • 0-20 mA
  • 4-20 mA

Power

  • 100-240 Vac, 50/60 Hz

Range

  • 0 to 1,000 μg/L
  • 0 to 10 mg/L

Response Time

  • t90 = 30 seconds

Temperature Range

  • Sample Temperature: 0° to 60° C (32° to 140° F)

Brand-Managed Distributed Enterprise Catalogs (DEC)
for Authorized Channel Partners
Book a DEC Demo
show more